Products/PRODUCTS CENTER
Product Classification/PRODUCT CATEGORIES
We focus on nuclear technology applications. Industry experts with decades of experience are the company's wealth, relying on their knowledge and rich experience, the company has been able to grow rapidly.
CT ionization chamber
The cylindrical pen-type CT ionization chamber can be used in conjunction with the CT dose phantom for measuring the dose-length product technical parameters of the CT machine.
Classification:
Products
Nuclear detector
Product accessories:

24-hour free hotline
Product Description
The cylindrical pen-type CT ionization chamber can be used in conjunction with the CT dose phantom for measuring the dose-length product technical parameters of the CT machine.
The ionization chamber was 10cm in length and had a good consistent response over the entire measurement length. Typical measurements are CTDI1 and CTDIW2, the dose-length product, which meets the requirements of IEC 61223-2-6.
Technical indicators
Product type: normal pressure pen type ionization chamber, in line with IEC 61223-2-6 medical CT imaging evaluation and routine test requirements.
Ionation chamber wall: PMMA coated graphite wall thickness: 1mm
Collection pole: aluminum tube outer diameter: 3mm, inner diameter: 2mm
Application: CT machine dose measurement
Measurement quantity: air kerma energy length product, exposure length product
Reference radiation quality: 120 kV, HVL 5.00mm Al (RQR9 specification)
Sensitive volume: 3.14 ㎝ 3
Sensitivity: 14nC/Gy
Long-term stability: ≤ ± 1.0%/year
X energy response: ≤ ± 4.0%
Leakage current: ≤ ± 10fA
Ionation chamber working voltage: ± 100-400v
Energy range: (70-150)kV
Irradiation field: (2*2) cm 2... (40*40) cm 2
The ionization chamber plug can choose BNC, TNC, etc., and can be fully used with PTW, IBA and other dosimeters.